Prospective assessment and validation of rhythmic and periodic pattern detection in NeuroTrend: A new approach for screening continuous EEG in the intensive care unit
Sprache des Titels:
Deutsch
Original Kurzfassung:
BACKGROUND:
NeuroTrend is a computational method that analyzes long-term scalp EEGs in the ICU according to ACNS standardized critical care EEG terminology (CCET) including electrographic seizures. At present, it attempts to become a screening aid for continuous EEG (cEEG) recordings in the ICU to facilitate the review process and optimize resources.
METHODS:
A prospective multicenter study was performed in two neurological ICUs including 68 patients who were subjected to video-cEEG. Two reviewers independently annotated the first minute of each hour in the cEEG according to CCET. These segments were also screened for faster patterns with frequencies higher than 4 Hz. The matching annotations (2911 segments) were then used as gold standard condition to test sensitivity and specificity of the rhythmic and periodic pattern detection of NeuroTrend.
RESULTS:
Interrater agreement showed substantial agreement for localization (main term 1) and pattern type (main term 2) of the CCET. The overall detection sensitivity of NeuroTrend was 94% with high detection rates for periodic discharges (PD = 80%) and rhythmic delta activity (RDA = 82%). Overall specificity was moderate (67%) mainly because of false positive detections of RDA in cases of general slowing. In contrast, a detection specificity of 88% for PDs was reached. Localization revealed only a slight agreement between reviewers and NeuroTrend.
CONCLUSIONS:
NeuroTrend might be a suitable screening tool for cEEG in the ICU and has the potential to raise efficiency of long-term EEG monitoring in the ICU. At this stage, pattern localization and differentiation between RDA and general slowing need improvement. This article is part of a Special Issue entitled "Status Epilepticus".